Agilent Technologies Helium G8610 Series Field Installation Instructions page 28

Ion source replacement for leak detection
Hide thumbs Also See for Helium G8610 Series:
Table of Contents

Advertisement

Vacuum Products Division
19. Connect the collimator wire to the collimator plate using an M 1.5 screw. See Figure 3-8.
20. Connect the 4 coupler screws to their respective filaments.
21. Insert the high voltage ion source wire (brownish-orange wire in rendering) into the ion chamber
and tighten the M1.6 socket head screw.
NOTE
Manual 699910005, Hot Ion Source Replacement
Collimator
screw
Figure 3-8. Collimator wire connection
Pull lightly with small pliers on the high voltage ion source wire
(brownish-orange) to ensure it is connected securely and will not pull out.
Figure 3-9. Ion chamber and extractor plate
Extractor
plate
Ion
chamber
Page 23