Agilent Technologies MSD 5977B Series Operating Manual page 70

Table of Contents

Advertisement

3
Installing Intuvo 9000 GC Columns
To Replace the 9000 GC/MS Tail
12 Remove the 9000 GC/MS Tail from the transfer line and bus. (See
Ferrule release hole
Transfer line
Bus click and run connection
Figure 15.
Exit in ion source
Figure 16.
13 Replace the gasket. (See
14 Carefully slide the 9000 GC/MS Tail into the GC/MSD interface, and gently
15 Check that the 9000 GC/MS Tail click and run fitting sits flat against the
16 Screw the 9000 GC/MS Tail column nut to the transfer line threaded
17 Loosely install the two compression bolts.
70
and
Figure
16.)
If the ferrule becomes stuck, press a pointed object, like a paper clip, into the
ferrule release hole at the end of the transfer line.
MS tail heater assembly clamp open with 9000 GC/MS Tail removed
9000 GC/MS Tail removed from GC/MSD Interface
page 63.)
place the click and run connection into the bus connection.
gasket in the bus connection.
connector finger tight, then use a ¼-inch wrench to tighten the nut an
additional 20 to 30 degrees.
Ferrule and nut
Click and run connection
"To Replace an Intuvo 9000 GC Gasket"
5977B Series MSD Operating Manual
Figure 15
on