8
CI Maintenance
To Connect/Disconnect Wiring to the EI HES Radiator
Entrance lens (1)*
Ion focus lens (2)*
Post extractor lens 2 (3)*
Post extractor lens 1 (4)*
Extractor lens (5)*
Ground wire
Heater wires
RTD wires
* The number in parenthesis is the
number engraved on the lens.
Figure 73
Wiring between the ceramic source board/feedthrough connector and the EI HES radiator
226
Ceramic source board
5977B Series MSD Operating Manual