4
Installing 7890 GC Columns
To Install the GC/MSD Interface Tip Seal
Interface tip seal
Knurled tip seal retainer
Tip seal base
Figure 24.
Forcing the analyzer closed if these parts are misaligned will damage the seal
C AU T I O N
or the interface or the ion source, or keep the sideplate from sealing.
4 Gently check the alignment of the analyzer and the interface.
5 You can align the analyzer and interface by wiggling the side plate on its
94
GC/MSD interface tip seal
When the analyzer is aligned correctly, the analyzer can be closed all the
way with no resistance except the spring tension from the interface tip seal.
hinge. If the analyzer still will not close, contact your Agilent Technologies
service representative.
5977B Series MSD Operating Manual