Installing 8890 GC Columns
To Install a Capillary Column in a Split/Splitless Inlet
Procedure
1 Cool the oven and inlet to room temperature.
2 Wearing clean gloves, press the column through the septum (this takes a bit
Capillary column
Column cutter
Ferrule, taper up
Inlet column nut
Septum
Figure 3.
3 Use the column cutter to score the column 2 cm or more from the end.
4 While holding the column, break the column end off at the score.
5 Inspect the end for jagged edges or burrs. If the break is not clean and even,
6 Wipe the outside of the free end of the column with a lint-free cloth
7 Position the column so it extends 4 to 6 mm past the end of the ferrule. (See
5977B Series MSD Operating Manual
of pressure). Then slide the column nut and conditioned ferrule onto the free
end of the column. (See
away from the column nut.
Preparing a capillary column for installation
repeat steps 3 and 4.
moistened with methanol.
Figure
4.)
Figure
3.) The tapered end of the ferrule should point
39