General Maintenance
Before Starting
Keep a record of system performance (tune reports) and maintenance
operations performed. This makes it easier to identify variations from normal
operation and to take corrective action.
Tools, spare parts, and supplies
Some of the required tools, spare parts, and supplies are included in the GC
shipping kit, MSD shipping kit, or MSD tool kit. You must supply others yourself.
Each maintenance procedure includes a list of the materials required for that
procedure.
High voltage precautions
Whenever the MSD is plugged in, even if the power switch is off, potentially
dangerous voltage (120 VAC or 200/240 VAC) exists on:
When the power switch is on, potentially dangerous voltages exist on:
Normally, all of these parts are shielded by safety covers. As long as the safety
covers are in place, it should be difficult to accidentally make contact with
dangerous voltages.
Perform no maintenance with the MSD turned on or plugged into its power
WAR N IN G
source unless you are instructed to by one of the procedures in this chapter.
Some procedures in this chapter require access to the inside of the MSD while
the power switch is on. Do not remove any of the electronics safety covers in any
of these procedures. To reduce the risk of electric shock, follow the procedures
carefully.
5977B Series MSD Operating Manual
The wiring and fuses between where the power cord enters the instrument
and the power switch
Electronic circuit boards
Toroidal transformer
Wires and cables between these boards
Wires and cables between these boards and the connectors on the back
panel of the MSD
Some connectors on the back panel (for example, the foreline power
receptacle)
169