4
Installing 7890 GC Columns
To Install a Capillary Column in the GC/MSD Interface Using a Standard Column Nut
17 You can align the ion source and interface tip seal by wiggling the side plate
18 Close the analyzer chamber door. (See
92
on its hinge. If the door still will not close, contact your Agilent Technologies
service representative.
page 220.)
"To Close the Analyzer Chamber"
5977B Series MSD Operating Manual
on