7
General Maintenance
To Assemble the EI HES
Figure 60
5 Place the heater/sensor assembly over the guide pins on the source body
6 Place the repeller onto the heater/sensor assembly with the flat side of the
7 Place the source mount over the repeller.
8 Finger tighten the two gold plated screws using a T6 Torx screwdriver to
Do not overtighten the screws in the source body. Damage may occur to the
C AU T I O N
repeller.
9 On the side of the filament block opposite the finger grip, orient the dual
10 Inset the filament block into the source holder, and use a T6 Torx screwdriver
5977B Series MSD Operating Manual
Secure the lens screw and lock ring insulator
with the four electrical pins oriented down the flat side of the source body.
repeller circumference aligned with the interface socket in the source body.
secure the source mount to the source body.
filament ceramic holder so that it lines up with the flat of the filament block.
Fully insert the three filament leads into the filament block.
to secure it to the holder with the gold plated screw. The filament block
should lay flat on the source holder. If it does not, remove the filament block
and make sure the filament is properly installed before reinserting the
filament block.