7
General Maintenance
To Clean an EI XTR, SS, or Inert Source
3 Collect the following parts to be cleaned for an EI XTR source: (See
These are the parts that contact the sample or ion beam. The other parts
normally should not require cleaning.
If insulators are dirty, clean them with a cotton swab dampened with
C AU T I O N
reagent-grade methanol. If that does not clean the insulators, replace them. Do
not abrasively or ultrasonically clean the insulators.
5977B Series MSD Operating Manual
on page 206.)
Repeller
Repeller block insert
Source body
Extractor lens
Ion focus lens
Entrance lens
Figure 66
205