Installing Intuvo 9000 GC Columns
To Condition an Intuvo Capillary Column
6 Program the column temperature to increase from 120 °C to the maximum
Never exceed the maximum column temperature, in the GC/MSD interface,
C AU T I O N
the GC oven, or the inlet.
7 Hold at the maximum temperature for 30 minutes.
The column is conditioned and ready to be used with your method.
5977B Series MSD Operating Manual
temperature limit for the column at a rate of 10 to 15 °C/min.