Agilent Technologies MSD 5977B Series Operating Manual page 121

Table of Contents

Advertisement

5
Operating in EI Mode
To Set Monitors for MSD Temperature and Vacuum Status
5 To set a monitor's alarm, double-click a monitor displayed in the Instrument
Figure 36
6 To make the new settings part of the method, save the Method.
5977B Series MSD Operating Manual
Control view to open that monitor's dialog for setting alarms. (See
Monitor's alarm settings
a Select the Set Alarm check box.
b Set the Warning Level, Alarm Level, and Below Minimum to appropriate
values.
c Enter descriptive text in the Monitor Label field if the default label is not
appropriate.
d Click OK to finish the monitor's alarm configuration.
Figure
36.)
121