Sampling Positions Dialog

The Sampling Positions dialog lets you position the logic analyzer's
setup/hold window (or sampling position) so that data on high-speed
buses is captured accurately, in other words, so that data is sampled
when it is valid.
When the device under test's data valid window is less than 2.5 ns
(roughly, for clock speeds >= 200 MHz), it's easiest to use eye finder
to locate the stable and transitioning regions of signals and to
automatically adjust sampling positions.
When the device under test's data valid window is greater than 2.5 ns
(roughly, for clock speeds < 200 MHz), it's easiest to adjust the
sampling position manually, without using the logic analyzer to locate
the stable and transitioning regions of signals.