Agilent Technologies 16718A Help Volume page 197

Logic analyzer
Table of Contents

Advertisement

Chapter 4: Concepts
Understanding Logic Analyzer Triggering
time. Two sequence levels can never be used to specify two events that
happen simultaneously.
For example, consider the following trigger sequence:
1. If ADDR = 1000 then Go to 2
2. If DATA = 2000 then Trigger
If the following samples were acquired, the logic analyzer would trigger
on sample #7.
Sample #
ADDR
1
1000
2
1010
3
1020
4
1030
5
1040
6
1050
7
1060
Note that the logic analyzer will not trigger on Sample #1 because a
new sample is acquired between the time that the condition in
Sequence level 1 is met and when the condition in Sequence Level #2 is
tested. A good way to think of this trigger sequence is "Find ADDR =
1000 followed by DATA = 2000 and then trigger". Multiple sequence
levels in a trigger sequence imply a "followed by".
Once a logic analyzer triggers, it does not trigger again. In other words,
even if more than one sample meets the trigger condition, the logic
analyzer still only triggers once. For example, using "ADDR=1000" as
our trigger, if the logic analyzer acquires the following samples, it will
trigger on Sample #2 and only on Sample #2.
Sample #
ADDR
1
0000
2
1000
3
2000
4
1000
5
1040
A frequently asked question is "What happens if the conditions in a
sequence level are not met?" For example, if there is a condition that
says "If ADDR = 1000 Then Trigger", what happens if the current
sample has ADDR = 2000? The logic analyzer simply acquires the next
sample and tries to execute this sequence level again. In essence, if the
trigger condition is "ADDR = 1000", this is equivalent to "Keep
acquiring more samples until you find one that has ADDR=1000".
Therefore, if you set up a trigger condition that is never met, the logic
196
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
DATA
2000
<- This sample meets the condition in Sequence level #1
3000
4000
5000
6000
7000
2000
<- This is where the logic analyzer triggers
<- The logic analyzer triggers here
<- The logic analyzer does NOT trigger again here

Advertisement

Table of Contents
loading

This manual is also suitable for:

16716a

Table of Contents