To stop a measurement
1. Select the Stop
If nothing happens when you start a measurement
• Analyzers with deep memory take a noticeable amount of time to complete
a run. Because data is not displayed until acquisition completes, it may
look like nothing is happening. Check the Run Status window to see if the
logic analyzer is still running.
• Messages such as "Waiting in level 1" may indicate you need to refine your
trigger.
• If the status shows as "Stopped", the analyzer either finished the
acquisition, or was unable to run. The cause of the problem is listed in the
bottom half of the Run Status window.
• Look for an error message in the message bar at the top of the window.
Common messages are "slow or missing clock" and "Waiting for trigger".
• If Run briefly changed to Stop or Cancel, select the Window menu,
choose the logic analyzer's slot, then choose the Waveform or Listing
display.
See Also
"Slow or Missing Clock" on page 180
"Waiting for Trigger" on page 184
"Error Messages" on page 170
To view the trigger status
While a logic analyzer measurement is running, you can view the
trigger status to see the sequence level that is evaluating captured
data, occurrence and global counter values, and flag values.
1. In the Trigger tab, select the Status subtab.
See Also
Run status (in the system help volume).
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
To select transitional timing or store qualified
starts all instruments currently placed in the workspace.
or Stop All button.
Chapter 2: Task Guide
87

Advertisement

Table of Contents
loading

This manual is also suitable for:

16716a

Table of Contents