Periodic Maintenance - Agilent Technologies 8860 User Manual

Gas chromatograph gasifier
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

6
Maintenance and Troubleshooting

Periodic Maintenance

Periodic Maintenance
This section contains some suggestions for ensuring good performance of the gasifier. The
maintenance interval varies with the use of the instrument.
User should not open the heating box to adjust or maintain the regulator or other parts.
WAR N IN G
On an occasional basis:
Check the gasifier vent/output flow for a possible leak (see
(see
If filter is used, replace it regularly
Make sure the mounting bracket screws are tight.
Be sure all cables are securely connected.
Calibrate the gasifier. The consistent output pressure is critical for reproducible analysis and
accurate quantitation. If your analysis needs be done regularly, it is recommended to verify the
system by using your calibration standard weekly to establish the experimental relative
response factor set. The results should be consistent, otherwise it may indicate there is
pressure drifting and/or your gasifier has leak or block issue. If your test is only done
occasionally, analyze the calibration standard before each set of analysis
The GC tracks gasifier usage. To view the current statistics, go to the Maintenance tab in the
Browser Interface (or GC touchscreen for 8890).
Maintenance related items like Early Maintenance Feedback (EMF) and specific GC Gasifier
hardware items with dedicated procedures will consist of the following, located under
Maintenance > GC Gasifier
Vaporizer
Transfer Line
46
Checking for
Blockage).
Agilent 8860, 8890, and 990 Micro GC Gasifier User Manual
Checking for
Leaks) or block

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

88908990 micro

Table of Contents