Installation for 8860 GC and 8890 GC

Connect the sample cylinder

Figure 5.
Connect the sample cylinder
1 Connect the LPG or high-pressure sample to inlet 1 or 2 that has a filter unit installed using a
5/16 inch wrench.
2 Connect a long vent line to the VENT outlet to safely guide hazardous fumes to a fume hood
or other appropriate vent.

Connect the power cord and USB cable

1 Lock the Core Ferrite (P/N 9170-2626) to the USB cable, about 50 cm away the USB
connector, Gasifier side.
2 Connect the gasifier USB cable to the side of the gasifier and to either USB1 or USB2 on
the GC.
3 Insert the USB cable and power cable into the two cable clamps and lock the clamps.
Agilent 8860, 8890, and 990 Micro GC Gasifier User Manual
Gasifier External Filter Assembly Structure