Installation for 8860 GC and 8890 GC
Connect the sample cylinder
Figure 5.
Connect the sample cylinder
1 Connect the LPG or high-pressure sample to inlet 1 or 2 that has a filter unit installed using a
5/16 inch wrench.
2 Connect a long vent line to the VENT outlet to safely guide hazardous fumes to a fume hood
or other appropriate vent.
Connect the power cord and USB cable
1 Lock the Core Ferrite (P/N 9170-2626) to the USB cable, about 50 cm away the USB
connector, Gasifier side.
2 Connect the gasifier USB cable to the side of the gasifier and to either USB1 or USB2 on
the GC.
3 Insert the USB cable and power cable into the two cable clamps and lock the clamps.
Agilent 8860, 8890, and 990 Micro GC Gasifier User Manual