Connect The Sample Cylinder - Agilent Technologies 8860 User Manual

Gas chromatograph gasifier
Hide thumbs Also See for 8860:
Table of Contents

Advertisement

4
Installation for 990 Micro GC

Connect the sample cylinder

Figure 8.
Connect the sample cylinder
1 Connect the LPG or high-pressure sample to inlet 1 or 2 socket using a 5/16 inch wrench.
2 Connect a long vent line to the VENT outlet to safely guide hazardous fumes to a fume hood
or other appropriate vent.
3 After all connections are completed, wait for the temperature setting to be ready. The process
takes approximately 40 minutes for the vaporizer to stabilize at 150°C from room
temperature, and approximately 15 minutes for the transferline to stabilize at 100 °C from
room temperature.
Agilent 8860, 8890, and 990 Micro GC Gasifier User Manual