Safety Information
General safety precautions
Substituting parts or performing any unauthorized modification to the Power Supply may
result in a safety hazard.
Before plugging in your instrument, verify that the voltage of your local power source is set
appropriately.
Turn off and cool down the GC and Gasifier before disconnecting the Transferline and GC.
Do not use an Uninterruptable Power Supply (UPS) with a Gasifier.
The input voltage of external power supply is 100-240V. Never use the mains supply which
voltage fluctuations are exceeded +/- 10 % out of this range.
Stop using Gasifier with frayed or damaged Transferline, and contact your local Agilent sales
office.
Gasifier will be turned on as long as the power cord is connected to the power source. There is
C AU T I O N
no "On/Off" switch.
Make sure that the sampler has been removed prior to maintaining, moving or relocating the
C AU T I O N
Gasifier. Prior to moving, all covers and enclosures must also be seated properly. Make sure all
covers and enclosures are secure.
Appliance inlet coupler (main input power cord) is the power disconnect device. Do not position
WAR N IN G
the instrument such that access to the coupler or plug is impaired.
Transferline could be operated at temperatures high enough to cause serious burns.
WAR N IN G
You should always cool the Gasifier to room temperature before working on it. If you must
perform maintenance on the transferline, use a wrench and wear thermally protective gloves.
Add additional Warning information
Do not place the Power Supply and Power cord on the top of GC or Gasifier when GC or Gasifier
WAR N IN G
is powered on.
Agilent 8860, 8890, and 990 Micro GC Gasifier User Manual