Note: the time parameters must meet the following conditions:
Step Number: it selects the number of frequency steps of the Upstair Sweep mode.
Start Frequency [Hz]: it selects the initial sweep frequency.
Stop Frequency [Hz]: it selects the final sweep frequency.
Note: The Start Frequency [Hz] and the Stop Frequency [Hz] are defined in the Carrier tab.

Sweep Trigger Mode

Repeat: the instrument starts when the Run/Stop button is pressed and repeats the sweep
continuously.
Trigger: When the Run/Stop button is pressed the instrument waits for a Trigger signal. When
the trigger is detected it generates the sweep profile then stops waiting for a new Trigger.
During the wait for a Trigger the instruments generates the waveform with its frequency equal
to the start frequency.

Burst

In Burst mode a waveform is repeated a predefined number of times or until the Trigger signal is at High
Level depending on the selected Burst Type. This Run Mode is available for all waveforms except the DC
level.
�������� �������� + �������� + ������������ ≤ 2000 ��