channel 2's frequency to maintain the specified ratio and offset in real time: if you set Channel 1
frequency to 1 MHz you will get 3.5 MHz on channel 2.
Example: Amplitude & Offset Coupling
1. Set the Channel 2 coupling button to ON and the coupling parameters will appear
2. Check the Amplitude and the Offset parameter
3. Set Amplitude Ratio to 1 and Offset to 0
4. Set Offset Ratio to 1 and Offset to 0
This coupling configuration ensures that the amplitude and offset parameters on both channels are the
same: if you move the amplitude or offset parameter on Channel 1 you will see that the amplitude and
offset on Channel 2 will change at the same time of the same quantity.

Certifications

Teledyne LeCroy certifies compliance to the following standards as of the time of publication. Please see
the EC Declaration of Conformity document shipped with your product for current certifications.

EMC Compliance

EC DECLARATION OF CONFORMITY - EMC
The instrument meets intent of EC Directive 2014/30/EU for Electromagnetic Compatibility. Compliance
was demonstrated to the following specifications listed in the Official Journal of the European
Communities:
EN 61326-1:2013, EN 61326-2-1:2013 EMC requirements for electrical equipment for measurement,
control, and laboratory use.
Electromagnetic Emissions:
EN 55011:2010, Radiated and Conducted Emissions Group 1, Class A
EN 61000-3-2/A2:2009 Harmonic Current Emissions, Class A
EN 61000-3-3:2008 Voltage Fluctuations and Flickers, Pst = 1
Electromagnetic Immunity:
EN 61000-4-2:2009 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling
4
planes
EN 61000-4-3/A2:2010 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2
GHz; 1 V/m, 2 GHz - 2.7 GHz
EN 61000-4-4/A1:2010 Electrical Fast Transient/Burst, 1 kV on power supply lines, 0.5 kV on I/O signal
4
data and control lines
EN 61000-4-5:2006 Power Line Surge, 1 kV AC Mains, L-N, L-PE, N-PE
EN 61000-4-6:2009 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz - 80 MHz
EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles, 0%/250 cycles