Teledyne TED 60-T Operator's Manual

Portable oxygen analyzer

Advertisement

TYPE B EQUIPMENT: Equipment providing a particular degree of protec-
tion against electric shock, particularly regarding—
allowable LEAKAGE CURRENT
Reliability of the protective earth connection
(if present).
Teledyne Analytical Instruments
P/N M40575
12/08/99
ECO # 99-0480
i

Advertisement

Table of Contents
loading

Summary of Contents for Teledyne TED 60-T

  • Page 1 TYPE B EQUIPMENT: Equipment providing a particular degree of protec- tion against electric shock, particularly regarding— • allowable LEAKAGE CURRENT • Reliability of the protective earth connection (if present). P/N M40575 12/08/99 ECO # 99-0480 Teledyne Analytical Instruments...
  • Page 2 Teledyne Electronic Technologies, Sensor Technologies, 16830 Chestnut Street, City of Industry, CA 91749–1580.
  • Page 3 These chapters help you to become better acquainted with the monitor and how it works before you actually begin to use it. Please read Chapter 2: Operations Chapter 2: Operations Chapter 2: Operations Chapter 2: Operations Chapter 2: Operations in its entirety before proceeding further with the manual. Teledyne Analytical Instruments...
  • Page 4 Teledyne warrants that the goods are free from defects of material and of constuction for a period of 2 years from the date of shipment from Teledyne. The Class T-7 Micro- Fuel Cell is warranted for one year from the date of shipment from Teledyne. The...
  • Page 5: Table Of Contents

    3 3 3 3 3 Maintenance and Troubleshooting Troubleshooting Table .................. 3-1 4 4 4 4 4 Appendix Specifications ....................4-1 Spare Parts List ..................... 4-2 Optional Accessories..................4-2 Repair Service ....................4-2 Material Safety Data Sheet ................4-3 Teledyne Analytical Instruments...
  • Page 6 Teledyne Analytical Instruments...
  • Page 7: General Information

    General Information General Information General Information The Teledyne Electronic Technologies 60T Oxygen Analyzer provides analysis in respirators, incubators, and other medical equipment where specific monitoring of oxygen is vital. The unit is compact, light, fits comfortably in a pocket, and incorporates a belt clip for easy carrying.
  • Page 8: Applications

    RFI Immunity RFI Immunity The Teledyne Model TED 60-T has been tested to the standard required by the Euro- pean Community for CE marking, EN 60601-1-2. The sample subjected to testing passed all sections of this testing with the exception of the following: From 26MHz to 250MHz, the display was unstable.
  • Page 9: Operations 2

    3. Lift the battery from its compartment and unsnap it from the snap connector. 4. Snap the new battery to the snap conector. (Match up the male and female snaps. The battery snap connector will only accept a properly oriented battery. Do not try to force it. Teledyne Analytical Instruments...
  • Page 10: Sensor Installation Or Replacement

    Micro-Fuel Cell Flow Diverter Adapter Sensing Surface Teledyne Analytical Instruments...
  • Page 11: Calibration

    2.) In confined volumes such as incubators or tents. When measuring for oxygen in breathing circuits, the diverter must must must must must be used. The diverter should be screwed onto the threaded front end of the T-7 sensor. A tee adapter Teledyne Analytical Instruments...
  • Page 12: Sterilization

    T-7. Exposing the sensor to air or gases that do not contain anesthetic agents for a period of time equal to or greater than the exposure interval will eliminate the reading error in most cases. Teledyne Analytical Instruments...
  • Page 13: Pressure

    (especially if the sensor is fast enough to sense the changes, as is the T-7). In reality, the percentage of oxygen is not changing; it is the total pressure that is increas- ing, producing a corresponding increase in the partial pressure of oxygen. A hundred- Teledyne Analytical Instruments...
  • Page 14: Humidity

    (e.g., incubators). No adjustments should be made during this period (about 1 to 2 hours), since this error will be eliminated when both the thermistor and sensing electrode have had sufficient time to come to thermal equilibrium. Teledyne Analytical Instruments...
  • Page 15: Discrepancy In Readings

    The information obtained from the TED 60T should never be used to make adjustments to the primary life-support system, but should only be used as an indication that the primary device may may require service and/or calibration. Teledyne Analytical Instruments...
  • Page 16: Do's & Don'ts

    Use cleaning agents or liquids in the cable receptacles or around the battery compartment. • Place the TED 60T unit itself in a water vapor-saturated environment. • Expose the LCD to excessive sunlight. • Expose the TED 60T to a condensing water environment such as a mist tent. Teledyne Analytical Instruments...
  • Page 17: Troubleshooting Table

    Humidity. ing absorbent tissue or cot- ton swab, gently wipe off sensing surface inside threaded portion of sensor assembly. B) Radio frequency interfer- B) Relocate unit away from ence (RFI). sources of RFI. Teledyne Analytical Instruments...
  • Page 18: Troubleshooting

    In the event that none of these procedures produce desired results, remove the In the event that none of these procedures produce desired results, remove the batteries and return the unit to Teledyne for repair. batteries and return the unit to Teledyne for repair.
  • Page 19: Appendix

    Stability: Less than 1% O over an 8 hour period at constant pressure and Stability: Stability: temperature. Warm up time: Warm up time: Warm up time: typically 20 minutes or less Warm up time: Warm up time: Teledyne Analytical Instruments...
  • Page 20: Spare Parts List

    Instruments and sensors damaged by accident or misuse are not covered by the Instruments and sensors damaged by accident Instruments and sensors damaged by accident Instruments and sensors damaged by accident warranty. In these situations, service charges will be based on time and materials. Teledyne Analytical Instruments...
  • Page 21 The MSDS on this material is available upon request The MSDS on this material is available upon request The MSDS on this material is available upon request through the Teledyne Environmental, Health and through the Teledyne Environmental, Health and through the Teledyne Environmental, Health and...
  • Page 22: Material Safety Data Sheet

    Micro-Fuel Cells Mini-Micro-Fuel Cells, all classes Super Cells, all classes except T–5x Oxygen Sensors, all classes. Manufacturer: Manufacturer: Manufacturer: Manufacturer: Manufacturer: Teledyne Electronic Technologies/Analytical Instruments Address: Address: Address: 16830 Chestnut Street, City of Industry, CA 91748 Address: Address: Phone: Phone:...
  • Page 23 OSHA: Not listed Other health hazards: Other health hazards: Other health hazards: Other health hazards: Other health hazards: Lead is listed as a chemical known to the State of California to cause birth defects or other reproduc- tive harm. Teledyne Analytical Instruments...
  • Page 24 TELEDYNE ELECTRONIC TECHNOLOGI,ES Sensor used only as a guide. TELEDYNE ELECTRONIC TECHNOLOGI,ES Sensor used only as a guide.
  • Page 25 Appendices 4 Appendices 4 Appendices 4 Appendices 4 Appendices 4 Teledyne Analytical Instruments...
  • Page 26 MONITEUR D’OXYGENE TED 60T Le moniteur d’oxygène modèle TED 60T fournit une CONSEILS D’INSTALLATION analyse pour les respirateurs, incubateurs, ainsi que les Comme c’est le cas pour tous les détecteurs d’oxygène, autres équipements médicaux pour lesquels le contrôle une condensation excessive sur la surface de détection spécifique d’oxygène est vital.
  • Page 27 à un usage impropre ou négligent, en cas Gamme des températures d’accident ou de réparations autres que celles exécutées d’opération: 32-104°F (0-40°C) par Teledyne ou le centre de service autorisé. Gamme des températures de stockage: 32-122°F (0-50°C) Temps de réponse: 90% en moins de 6 secondes à...
  • Page 28 MONITOR PER OSSIGENO TED 60T Il monitor per ossigeno modello TED 60T fornisce un’analisi CONSIGLI PER L’INSTALLAZIONE DEL MONITOR TED durante l’impiego di respiratori, incubatrici ed altri strumenti medici in cui il controllo dell’ossigeno è di vitale importanza. Come in tutti i rilevatori di ossigeno un eccessiva condensa Lo strumento è...
  • Page 29 SPECIFICHE che di difetti del materiale impiegato per due anni a partire Range: 0-100% Ossigeno dalla data di uscita dell’articolo dalla sede Teledyne, ad Precisione: ± 2% su scala completa a esclusione della micro componente di alimentazione temperatura costante, ± 5% su...
  • Page 30 MONITOR DE OXIGENO TED 60T El Monitor de Oxigeno Modelo TED 60T provee análisis SUGERENCIAS PARA LA INSTALACION en los respiradores, incubadoras y demás equipo médico Como en todo los sensores de oxigeno, la excesiva en que es vital la vigilancia específica del oxigeno. El equipo condensación sobre la superficie sensible bloqueará...
  • Page 31 Teledyne, excepto delantera y trasera del conjunto del sensor están en el caso de la Pila Micro-Fuel T-7 Clase T, para la que se a diferentes temperaturas.
  • Page 32 TED 60T OXYGEN MONITOR (SAUERSTOFF-KONTROLLGERÄT) Das Modell TED 60T Oxygen Monitor ermöglicht Analysen Deckels, so daß Sie Zugang zur 9-Volt-Batterie für Atemgeräte, Inkubatoren und andere medizinische erhalten. Geräte, bei denen bestimmte Sauerstoffkontrollen Entfernen Sie die verbrauchte Batterie und lebenswichtig sind. Das Gerät ist kompakt, leicht, paßt ersetzen Sie sie durch eine neue 9-Volt- einfach in eine Hosen- oder Jackentasche und kann durch Alkalibatterie.
  • Page 33 Minuten lang mit Wasser ausgespült werden und GARANTIE ärztliche Hilfe gerufen werden. TED stellt Teledyne übernimmt die Garantie dafür, daß die Produkte sogenannte Material Safety Data Sheets (MSDS hinsichtlich ihres Materials und ihrer Herstellung für einen - Angaben zur Sicherheit des Materials) bereit.

Table of Contents