Cetus
5.
Remove the electromagnetic brake protection
shroud.
6.
Remove the electromagnetic brake screws
(ø3 x 15L x 3pcs).
7.
Take out the electromagnetic brake.
Please refer back in reverse for reassembly and
replacement.
Caution:
When reassembling the electromagnetic brake,
please pay attention to the position of the gear
lever.
Electromagnetic brakes shall be cleaned
regularly. After removing the electromagnetic
brake, remove the three hexagon socket screws
(pictured as below) on the electromagnetic
brake. After removing the brake pad, clean the
brake pad with a brush (do not mix with
detergent, just clean it with a dry brush).
S81201-92800~A
8-3
Differential mechanism
maintenance:
Hex wrench 13 x 2pcs
1.
Refer to 2-1 Battery power-off: The scooter is
powered off.
2.
Refer to 6-4 Main controller maintenance:
Unplug the motor power line.
3.
Remove the motor mount screws (M8 x 45L x
4pcs) on both sides.
4.
The motor holder on both sides can be
removed.
5.
The entire set of differential mechanism can
be removed.
Please refer back in reverse for reassembly and
replacement.
Service Manual