Charger failure and troubleshooting
The power charging indicator light (red light) is not
on:
Troubleshooting → Confirm whether the socket
has normal voltage. If the socket voltage is normal,
please replace with a new charger directly.
The charging indicator light (orange light) is not on:
Troubleshooting → Confirm that the charging
output connector is properly connected to the
battery connector. If it is correct, it may be a battery
failure and a new charger should be replaced with.
The charging indicator light (orange light) changes
quickly (to green light):
Troubleshooting → Confirm that the battery is fully
charged. If it is not full, it may be a battery failure
and a new charger should be replaced with.
Caution:
Charging time is about 8 hours (depending on
battery discharge degree). Please do not
charge continuously for more than 15 hours.
Please charge in a well-ventilated place, and
avoid charging in outdoor or rainy environments.
S81201-92800~A
The display panel automatically switches to the
SPD speedometer mode while in the charging
state.
Display status as
shown on the
following page
The AC power plug must be plugged directly
into the AC socket on the wall. Do not use an
extension cable or adapter plug to avoid danger.
When connecting, please make sure that both
ends of the connectors are dry and clean.
Battery electricity table:
Battery residual display
Cell
number
status
display
After the power is turned on, an alarm will be set
off when the electric quantity is lower than 30%
(beeping for three times)
Charging indication table:
Electric
quantity
40 %
2 cells
Charging
Loop
indication
display
55 %
3 cells
Loop
display
Service Manual
Symbol flashing
Warning
indicator,
continuous
flashing
Cell number
status display