Cetus
70 %
4 cells
Loop
display
80 %
5 cells
Loop
Charging
display
90 %
indication
6 cells
Loop
display
100 %
6 cells
Constantly
on
Incremental
0.5 seconds
frequency
1. The state of the cell number
only increases upwards and
does not decrease.
2. The charger PIN3 (CH3) is
used as the determination
signal. When CH3 is grounded,
Action
it enters the charging mode,
and KEY ON or KEY OFF is
characteristics
not limited.
3. Press any key, backlight will
be turned on, and it will
automatically turn off if no
touching the button within five
seconds.
The cell number status display
Remarks
is for reference only, and the
exact judgment still depends
on the indicator light on the
charger.
S81201-92800~A