Turning Deceleration Set Maintenance - Invacare Cetus Service Manual

Hide thumbs Also See for Cetus:
Table of Contents

Advertisement

Cetus
5.
The rear compartment can be separated.
Please refer backward for reassembly and
replacement.
Caution:
In this step, only the back shroud of the
compartment can be removed. If it is needed to
remove the front shroud of the compartment,
please refer to 5-2 Maintenance of front shroud of
front compartment.
* Please take out the articles in the compartment before
4-3
Turning deceleration set
disassembling so as to avoid falling or losing.
maintenance:
Cross screwdriver x 1pcs
1.
Refer to 2-1 Battery power-off: The scooter is
powered off.
2.
Please refer to 4-1 Front slanting
maintenance: open the front slanting.
3.
Remove the turning deceleration set
connector.
S81201-92800~A
4.
Remove the locking screw (M5 x 13L x 2pcs).
5.
The turning deceleration set can be removed.
Please refer backward for reassembly and
replacement.
Caution:
The turning deceleration set is used to control
the turning and deceleration of the scooter. If
the turning and deceleration function are
abnormal, please check whether the micro
switch connector is falling off. If the connector is
normal, please replace with a new micro-switch.
Some scooters do not use the turning and
deceleration set. They may adopt the gyroscope
to achieve the function of turning and
deceleration. For details, please refer to:
4-5 Gyroscope Maintenance.
The arrow on the micro-switch box must be
aligned with the arrow on the micro switch
launch pad.