Temperature Sensor Maintenance; Gyroscope Maintenance - Invacare Cetus Service Manual

Hide thumbs Also See for Cetus:
Table of Contents

Advertisement

Cetus
4-4
Temperature sensor
maintenance:
Cross screwdriver x 1pcs
1.
Refer to 2-1 Battery power-off: The scooter is
powered off.
2.
Please refer to 4-1 Front slanting
maintenance: remove the front slanting.
3.
Remove the temperature sensor connector.
4.
The temperature sensor can be removed
from the slot.
Please refer backward for reassembly and
replacement.
Caution:
The temperature sensor is fixed with the slot of
the frame. Please fix it tightly when
reassembling so as to avoid falling off.
4-5

Gyroscope Maintenance:

Cross screwdriver x 1pcs
S81201-92800~A
1. Refer to 2-1 Battery power-off: The scooter is
powered off.
2. Please refer to 4-1 Front slanting maintenance:
open the front slanting.
3. Remove the gyroscope connector.
4. Remove the locking screw (M4 x 12L x 2pcs).
5. The gyroscope set can be removed.
Please refer backward for reassembly and
replacement.
Service Manual
18