COMMISSIONING

7.3.10.4 Configuration of Gauge Using SmartLink and Engauge

7.3.10.4.1 Configuration of SmartLink with TRL/2 card

7.3.10.4.2 Configuration of the Smartlink system

Part No.: 4417340 Revision 1
The smartlink system requires Engauge service tool for configuration.
The Engauge service tool (version 2.2 and higher) is required for
SmartLink modules delivered from September 2007 with software
versions:
HCM -GPU A1005 (and higher)
FCM-TRL/2
The Connection between the PC and the Smartlink
Before you can start to configure the SmartLink you have to connect
the SmartLink to PC with the service tool Engauge installed. The
communication between the PC and the SmartLink takes place by
means of a serial connection. The user must select a free serial port on
the PC. The chosen port must be set in Engauge, the maximum com-
port number is COM24. On the SmartLink side the communication
takes place by using the Non Isolated RS232 interface on the HCM-
GPU module.
Making new site in Engauge
NOTE: Navigate to C:\Users\Public\Documents\Honeywell\ConfigFiles
BoardDescriptor" and paste board descriptor files for FCM-TRL/2 and
HCI-TRl/2.
Service Manual
954 SmartServo FlexLine
165