COMMISSIONING
Name
Permitted Values

[Contact 2

Status
health]
GOOD
UNCERTAIN
(RO)
BAD
Status Category
GOOD, ACTUAL VALUE
GOOD, STORED VALUE
UNCERTAIN,
INSTRUMENT
BAD, HARDWARE
FAILURE
BAD, OPERATIONAL
FAILURE
Status Code
See section 7.3.13.1
"Safety Function Status"
for definitions.
[Contact 2
< Not Used >
safety function]
< Underfill >
< Overfill >
[Contact 2
< 0 – 3.402823 × 10
threshold]
(meters)
Part No.: 4417340 Revision 1
Contact 2
Default Value
< Overfill >
38
>
< 0 >
Service Manual
954 SmartServo FlexLine
Description
Functional health of contact 2.
Any status other than GOOD
will result in a safe state
activation for contact 2.
The condition of product level
relative to the threshold which
will determine the activation of
a contact 2 safe state, as an
underfill or overfill condition.
Overfill: activation occurs when
the product level is above the
Overfill Threshold.
Underfill: activation occurs
when the product level is
below the Underfill Threshold.
Not Used: contact 2 is
activated in a constant safe
state and does not perform
any function.
The product level limit value
for determining an overfill or
underfill condition for contact
2.