Agilent Technologies 81250 System Programming Manual And Scpi Reference page 222

Parallel bit error ratio tester
Hide thumbs Also See for 81250:
Table of Contents

Advertisement

DeMUX Rewiring Parameter Commands
Return Value
Parameters
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:<Handle>:SGENeral:GLOBal:
SYNChronization:RMODe?
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronizat:RMODe?
Returns the current mode for DeMUX rewiring. Can be:
• AUTO
The automatic mode tries first to rewire with the trace detection
method and, if that fails, with the terminal roundtrip method.
• TRO
(Terminal ROundtrip) Rewire takes place with the terminal
roundtrip.
• TDET
(Trace DETection) Rewire takes place with the trace detection
method.
Example
:_test:SGENeral:GLOBal:SYNCchronization:RMODe?
might return:
AUTO or TRO or TDET
:<Handle>:SGENeral:GLOBal:
SYNChronization:RCH
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronizat:RCHeck ON|OFF
<ON|OFF>
If verify rewiring is switched on, the system performs additional
inspections and complains about possible errors.
If verify rewiring is switched off, the system tries to ignore noncritical
problems to carry out a measurement.
Example
:_test:SGEN:GLOB:SYNC:RCH ON
or
:_test:SGEN:GLOB:SYNC:RCH OFF
Commands in DSR Application Subsystems
Switches the rewire verify mode on or off.
221