:SGENeral:GLOBal:TRIGger Subsystem
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Limits
Works only on a stopped machine, range from 1 ... 256.
Example
_TEST:SGEN:GLOB:TRIG:CLOC:DIV 2
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:DIVider?
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:DIVider?
Queries the current clock divider.
Example
:_test:sgen:glob:trig:cloc:div?
might return:
1
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:RCLock:DETect
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:RCLock:DETect
[ONCE]
This command measures the external clock reference and sets the
corresponding mode automatically. The automatically set modes are:
EXT1 reference
EXT2 reference
EXT5 reference
EXT10 reference
The E4809A can only use the EXT10 mode. The E4805B and E4808A
can use all modes.
Example
:_test:SGEN:GLOB:TRIG:RCL:DET
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLock:MEASurement
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLock:
MEASurement
This command measures the external clock and sets the
corresponding mode EXTernal automatically.
Commands in DSR Application Subsystems
internal clock source with 1 MHz external clock
internal clock source with 2 MHz external clock
internal clock source with 5 MHz external clock
internal clock source with 10 MHz external clock
177