Agilent Technologies 81250 System Programming Manual And Scpi Reference page 177

Parallel bit error ratio tester
Hide thumbs Also See for 81250:
Table of Contents

Advertisement

Commands in DSR Application Subsystems
Parameter
Return Value
Parameters
Range/Value Coupling
Reset Value
176
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Example
:_test:SGEN:GLOB:TRIG:CLOC:MEAS:DIR
or
:_test:SGEN:GLOB:TRIG:CLOC:MEAS
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:MULTiplier
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:MULTiplier
<Multiplier>
<Multiplier>
256.
This command works only when the Agilent 81250 has been stopped.
Example
:_test:SGEN:GLOB:TRIG:CLOC:MULT 10
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:MULTiplier?
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:
MULTiplier?
Returns the current clock multiplication factor.
This command works only when the Agilent 81250 has been stopped.
Example
:_test:SGEN:GLOB:TRIG:CLOC:MULT?
might return:
10
:<Handle>:SGENeral:GLOBal:TRIGger[:
SEQuence][:LAYer]:CLOCk:DIVider
Syntax
:<Handle>:SGENeral:GLOBal:TRIGger[:SEQuence][:LAYer]:CLOCk:DIVider
<DIVider>
<DIVider>
The clock divider.
Period, frequency, measured value.
1
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:SGENeral:GLOBal:TRIGger Subsystem
Sets the clock multiplication factor in a range of 1 ...