Commands in DSR Application Subsystems
Description
180
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
...:SOUR:STOPsignal
soon as a stop signal exceeds the input threshold. The sequence can be
stooped by a positive or negative input signal depending on the
parameter set.All these modes are active if the system is started
manually or started by a remote command.
Pulse stream
Input signal
Pulse stream
Input signal
NO TE
IMMediate and STARt are always supported, GATed and STOP are
only supported for E4832A modules.
:<Handle>:SGENeral:GLOBal:ARM[:
SEQuence][:LAYer][:SOURce]
Syntax
:<Handle>:SGENeral:GLOBal:ARM[:SEQuence][:LAYer][:SOURce] IMMediate |
GATed | STARtsignal | STOPsignal
This command controls the start mode of the Agilent 81250 System. If
IMMediate is selected, then the sequence can be started and stopped
by the :SGEN:GLOB:INIT:CONT ON | OFF command, or it is possible
to press the start or stop button in the graphical user interface.
NO TE
If the external clock mode is set and the external start mode is
selected, the external clock must be in a stable state for 100 ms before
the external start can take place.
Example
:_test:SGEN:GLOB:ARM IMM
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
:SGENeral:GLOBal:ARM Subsystem
The pulse stream (sequence) is stopped as