Agilent Technologies 81250 System Programming Manual And Scpi Reference page 106

Parallel bit error ratio tester
Hide thumbs Also See for 81250:
Table of Contents

Advertisement

:EDIT:SEGMent(*) Subsystem
Commands in DSR Application Subsystems
BIN Format
For BIN format, the returning stream will use the following format:
#abbbbccccccccccc where:
• a is the number of b's
• bbbb is the number of c's
• ccccccc is the raw data; in the GUI, this raw data is presented as
ASCII characters.
Return Value
Returns the data stream corresponding to the region specified in the
input.
Example
:_test:EDIT:SEGM3:PATT:DATA? 0,0,3,2,HEX
might return the following segment data:
A040A0E0
:_test:EDIT:SEGM3:PATT:DATA? 0,0,3,2,BIN
might return the following segment data:
#9000000004 @ à
The following figure shows how the hex characters correspond to the
segment data as shown in the user interface.
A 0
4
0
A 0 E 0
105
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006