Commands in DSR Application Subsystems
Parameters
220
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:<Handle>:SGENeral:GLOBal:
SYNChronization:RMODe
Syntax
:<Handle>:SGENeral:GLOBal:SYNChronizat:RMODe <AUTO|TRO|TDET>
<AUTO | TRO | TDET>
• AUTO
The automatic mode tries first to rewire with the trace detection
method and, if that fails, with the terminal roundtrip method.
• TRO
(Terminal ROundtrip) Rewire takes place with the terminal
roundtrip.
• TDET
(Trace DETection) Rewire takes place with the trace detection
method.
For more information on DeMUX rewiring modes, please refer to
"DEMUX Rewiring Modes" in the Agilent 81250 Parallel Bit Error
Ratio Tester System User Guide.
Example
:_test:SGENeral:GLOBal:SYNCchronization:RMODe AUTO
or
:_test:SGENeral:GLOBal:SYNCchronization:RMODe TRO
or
:_test:SGENeral:GLOBal:SYNCchronization:RMODe TDET
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
DeMUX Rewiring Parameter Commands
Sets the modes for DeMUX rewiring.