Agilent Technologies 81250 System Programming Manual And Scpi Reference page 110

Parallel bit error ratio tester
Hide thumbs Also See for 81250:
Table of Contents

Advertisement

:EDIT:SEGMent(*) Subsystem
Parameters
Return Value
Agilent 81250 Parallel Bit Error Ratio Tester, Programming Reference, March 2006
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
:<Handle>:EDIT:SEGMent(*):PATTern:FIND:
FIRSt?
Syntax
:<Handle>:EDIT:SEGMent*:PATTern:FIND:FIRSt? <data>, <VECT | TRAC>,
[NORMal | RVS]
<data>
Data sequence to be searched for. It can be one of the
following:
• Binary sequence (default, with optional "B" or "b" after the
sequence)
• Octal sequence (with mandatory "O" or "o" after the sequence)
• Hexadecimal sequence (with mandatory "H" or "h" after the
sequence)
<VECT | TRAC>
traces or vectors.
Width
Segment x
Length
Search Vectors
[NORMal | RVS]
NORMal.
• NORMal
The search is started at trace 0, vector 0, and progresses in the
incrementing trace, incrementing vector direction.
• RVS
The search is started at trace n–1, vector 0, and progresses in the
decrementing trace, incrementing vector direction.
d, v
Returns the digit index and vector index (comma-separated) of
the starting point of the first found instance of the data sequence.
Returns coordinates -1,-1 if the data sequence was not found.
Commands in DSR Application Subsystems
Defines if the data sequence is searched for in
Segment x
Search Traces
Defines if the direction of the search. Default is
109