Data Logging; A.9.11.2 Data Logging - Siemens SITRANS F Operating Instructions Manual

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Modbus
Data type /
register
Size in bytes
6182
Unsigned / 2
6099
Unsigned / 2
A.9.11.2

Data logging

Table A-75
Data logging
Modbus
Data type /
register
Size in bytes
11400
Unsigned / 2
11401
Unsigned / 2
11402
Unsigned / 4
FST020 IP65 NEMA 4X
Operating Instructions, 09/2017, A5E41425845-AB
Parameter
Description
Backup
Command to overwrite
the existing backup da‐
ta sets.
This ensures that the
current sensor serial
number and transmitter
serial numbers are up‐
dated in the backup da‐
taset. This enables the
automatic backup.
Restore
Command to restore
backup data. Coding is
bitwise.
Parameter
Description
Enable / disable Controls data logging.
Data logging
Controls data logging
mode
mode.
Logging interval Defines interval for log‐
ging. Granularity is 10
ms.
A.9 Modbus holding register tables
Default value
Value range /
[unit]
Setting options
(unit register)
-
● 1: Overwrite existing
backup files
-
● Bit 0: Restore sensor
setup parameters
● Bit 1: Restore sensor
application setup
parameters
● Bit 2: Restore
transmitter
application setup
parameters
● Bit 3: Restore
customer transmitter
configuration setup
parameters
● Bit 4: Restore
totalizers
● Bit 5: Restore system
setup
● Bit 6: Restore
customer sensor
configuration setup
parameters
Default value
Value range /
[unit]
Setting options
(unit register)
0
● 0: Disabled
● 1: Enabled
0
● 0: Log instantaneous
● 1: Log average value
500
1 - 8640000
[x 10 ms]
Modbus communication
Access
level
Write on‐
ly
Write on‐
ly
Access
level
Read /
write
Read /
write
Read /
write
161

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fst020 ip65 nema 4x

Table of Contents