Commissioning

6.5 Initial startup

6.2

General requirements

Before commissioning it must be checked that:
● The device has been installed and connected in accordance with the guidelines provided
in Installing/mounting (Page 19) and Connecting (Page 25).
6.3

Power-up

Power up the device. Local display will show a screen for initial startup (Page 36).
6.4

Local display

The device is commissioned/operated with the touch keypad on the local display.
The graphic display above the keypad gives a menu-guided operation of the individual device
function/parameters. Successful operation of the key is confirmed by tactile feedback as key
is pressed.
Figure 6-1
Note
HMI timeout
If no key is pressed for 10 minutes, the display switches to show operation view. If Backlight
is set to Automatic, display backlight goes off automatically 30 seconds after the last key press.
6.5
Initial startup
The first time the device is powered up, you will be prompted to set the language. The device
always starts up showing "Language" in English. When the language has been set, you will
be prompted to set the date and time.
Before using the flow meter for the first time, essential parameters should be considered. After
confirming/changing date and time you can choose to accept the default values or start the
Quick commissioning wizard.
36
Full graphical display
Touch keypad
Local display
FST020 IP65 NEMA 4X
Operating Instructions, 09/2017, A5E41425845-AB