Process Values; A.9.12.2 Process Values - Siemens SITRANS F Operating Instructions Manual

Ultrasonic flowmeters
Hide thumbs Also See for SITRANS F:
Table of Contents

Advertisement

Modbus communication
A.9 Modbus holding register tables
A.9.12.2

Process values

Simulate process values
Table A-80
Channel 2 output current
Modbus
Data type /
register
Size in bytes
2700
Unsigned / 4
2702
Float / 4
2704
Float / 4
2708
Float / 4
2738
Float / 4
2740
Float / 4
2742
Float / 4
164
Parameter
Description
Enable simula‐
Defines enabling or dis‐
tion
abling simulation of
multiple process values
at a time.
Density
Process density simula‐
tion value
Medium temper‐
Medium temperature
ature
simulation value
Viscosity
Process viscosity simu‐
lation value
Volume flow
Volume flow simulation
value
Sound velocity
Sound velocity simula‐
tion value
Flow velocity
Flow velocity simula‐
tion value
Default value
Value range /
[unit]
Setting options
(unit register)
0
Set bit means enabling:
● Bit 0: Density
● Bit 1: Medium
temperature
● Bit 3: Viscosity
● Bit 18: Volume flow
● Bit 19: Sound velocity
● Bit 20: Flow velocity
1000.0
0.08 to 10000.0
[kg/m
]
3
(7600)
20.0 [°C]
-50.0 to 250.0
0.000001
0.00000001 to 10.0
[m
/s]
2
(7524)
0.0
-3000.0 to +3000.0
[m
/s]
3
(7500)
1500.0
200.0 to 3000.0
[m/s]
(7648)
0.0
-40.0 to 40.0
[m/s]
(8014)
Operating Instructions, 09/2017, A5E41425845-AB
Access
level
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
Read /
write
FST020 IP65 NEMA 4X

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Fst020 ip65 nema 4x