Sagem Ultima 600 Maintenance Manual page 44

Table of Contents

Advertisement

S
• The unit performs a calibration (30'')
• Validate (key <>).
• Enter the values of the 2nd bottle specified
on the certificate joined with the bottle using
the arrow keys.
• Validate the function (key <>).
• Connect the quick coupling of the bottle
to the plastic connector at the rear of the
unit (See fig. 3.4).
• Open the gas bottle checking that
there is a slight overpressure.
• Wait for values on screen to stabilize.
• Validate using the <> key.
• Close the bottle.
• Disconnect the coupling.
• Validate using the <> key.
• Select the INFO ADJUSTMENT function
using the s or t key.
• Validate the function (key <>).
Issue 05/99
Ref.: 21821085-0/A
ULTIMA 600
Calibration in progress
Adjustment 2 points: Point #2
Target values:
CO= 3,49 xvol
HC= 2032 ppmvol (prop)
Open the bottle
CO= 0,00 xvol
CO2= 0,0 xvol
HC= 0000 ppmvol
<I>
Adjustment 2 points: Point #2
Adjustment in progress
<C>
30 s
CO2= 13,7 xvol
(3,49 xvol)
(13,7 xvol)
(1020 ppmvol)
5-8