Adjustement Information; Cell Status - Sagem Ultima 600 Maintenance Manual

Table of Contents

Advertisement

S
4.1.5 Adjustment information
Purpose:
By selecting the "Info Adjustment" function, you can read the date and the organization to
which the person who performed an adjustment belongs.

4.1.6 Cell status

Purpose:
By selecting the "IR Cell Status" function, you can read or print out a record of the statuses of
the IR cell.
• Select the "IR Cell Status" function
using the s or t key.
• Validate
the
• Display the statuses of the IR cell.
• Interpretation of statuses
#0=General status
0X0001= Heating in progress
0X0002= Calibration required
0X0004= New O
0X0008= Caution/Zero Calibr. (see status#1,#2)
0X0010= Bad O
0X0020= Pressure low
0X0040= New NOX sensor installed
0X0080= NOX sensor bad
0X0100= Caution: Condensation
0X4000= A/D function damaged
0X8000=WARNING internal cell
#1=Zero status
The following warning message is displayed when the zero level changes by more
than 5% with respect to the last zero.
0X0001= HC Zero Warning
0X0002= CO Zero Warning
0X0004= CO
2
0X0008= HiHC Zero Warning
0X0010= Reference Zero Warning
0X0020= O
Zero Warning
2
0X0040= No
Zero Warning
x
0X0080= RPM Zero Warning
Issue 05/99
function
(key
<>)
sensor installed
2
sensor
2
Zero Warning
Ref.: 21821085-0/A
Calibration
HC residue test
Info Adjustment
IR cell status
IR cell status SELFTEST
#0.4002 #1.0000 #2.0000 #3.0000
#4.0000 #5.0000 #6.0000 #7.0000
#8.0000 #9.0000 #10.0000
IMP
SELFTEST
4-5