Sagem Ultima 600 Maintenance Manual page 28

Table of Contents

Advertisement

S
#2= 1 point calibration status
The following warning message is displayed when the calibration level changes by
more than 5% with respect to the last calibration.
0X0001= HC Cal Warning
0X0002= CO Cal Warning
0X0004= CO
2
0X0008= HiHC Cal Warning
0X0010= Reference Cal Warning
0X0020= O
Cal Warning
2
0X0040= No
Cal Warning
x
0X0080= RPM Cal Warning
0X0100= PRESS Cal Warning
#3= 2 point calibration status
The following warning message is displayed when the calibration level changes by
more than 5% with respect to the last calibration.
0X0001= HC Cal Warning
0X0002= CO Cal Warning
0X0004= CO
2
#4= Status reserved for production
All bits must be at zero.
#5= Status reserved for production
All bits must be at zero
#6= Operational Warning cell
0X0002= Heating unit not controlled
0X0004= O
offset voltage Warning
2
0X0008= NO
X
0X0100= ROM Cheksum faulty on test after power up
0X2000= Brightness < 65%
0X8000= EEPROM incompatible
#7= A/D conversion circuit
0X0001= HC circuit fault
0X0002= CO circuit fault
0X0004= CO
2
0X0008= O
circuit fault
2
0X0010= No
circuit fault
x
0X0020= RPM circuit fault
0X0040= OIL circuit fault
0X0080= Ambient T circuit fault
0X0100= Pressure circuit fault
0X0200= Ref. circuit fault
0X0400= Transmitter fault
0X0800= Det circuit fault
#8= Calibration required
0X0080= O
required
2
0X0100= O
calibration required
2
0X0200= NO
X
0X0400= NO
X
#9 = Reserved
#10= approximate preheating time in seconds
Issue 05/99
Cal Warning
Cal Warning
offset voltage not within limits
circuit fault
zero required
calibration required
Ref. 21821085-0/A
4-6