Sagem Ultima 600 Maintenance Manual page 42

Table of Contents

Advertisement

S
• The unit performs a calibration.
• Standby screen displayed
Tightness test
Version, CS, FEP
Lambda control,
HC residue test
• Select the ADJUSTMENT 2 POINT function
using the s or t key.
• Validate the function (key <>).
• Enter the operator identification
#02 18/02/1999 is automatically displayed
? . . . = Dept. (ASD, for example) entered by
operator using keyboard.
st
? . . . . . . . . . = 1
first two letters of last name
of operator entered on keyboard.
• Validate the function (key <>).
• Enter the bottle values specified on the certificate
joined to the bottle using the arrow keys.
• Validate the function (key <>).
Issue 05/99
See
Normal
Service
Mode
letter of first name
Ref.: 21821085-0/A
ULTIMA 600
Calibration in progress
Tightness test
Version, CS,FEP
Lambda control
Calibration
HC residue test
Adjustm. 1 point
Adjustm. 2 points
Ajustage 2 points
Info adjustment
Adjustment 2 points: Point #1
Operator's identification
#02 18/02/1999 ? . . . ? . . . . . . . . .
Adjustment 2 points: Point #1
Target values:
CO= 3,49 xvol
HC= 2032 ppmvol (prop)
SELFTEST
SELFTEST
CO2= 13,7 xvol
5-6