Sagem Ultima 600 Maintenance Manual page 43

Table of Contents

Advertisement

S
• The unit performs a calibration (30'').
• Connect the quick coupling of the bottle to the
plastic connector at the rear of the unit
(See fig. 3.4).
• Open the gas bottle checking that
there is a slight overpressure.
• Wait for values on screen to stabilize.
• Validate using the <> key.
• Close the bottle.
• Validate using the <> key.
+
If scatter is too wide, following screen is displayed:
• Screen displayed
• Validate the function (key <>).
Issue 05/99
CO= 0,00 xvol
CO2= 0,0 xvol
HC= 0000 ppmvol
Adjustm. 2 points: Point #1
Check bottle values
then confirm or close
bottle and cancel
<I>
Target values:
CO= 3,49 xvol
HC= 2032 ppmvol (prop)
Ref.: 21821085-0/A
ULTIMA 600
Calibration in progress
Open the bottle
(3,49 xvol)
(13,7 xvol)
(1020 ppmvol)
<I>
Adjustment 2 points: Point #1
Adjustment in progress
<C>
Adjustment 2 points: Point #2
CO2= 13,7 xvol
30 s
5-7