Sagem Ultima 600 Maintenance Manual page 39

Table of Contents

Advertisement

S
• The unit performs a calibration (30'').
• Connect the quick coupling of the bottle
to the plastic connector on the rear side
of the unit (See fig. 3.4).
• Open the gas bottle, checking that there
is a slight overpressure.
• Wait for the values on the screen to stabilize.
• Validate using the <> key.
• Close the bottle.
+
If scatter is too wide, following screen is displayed:
• Select the INFO ADJUSTMENT function
using the s or t key.
• Validate the function (key <>).
Issue 05/99
Adjustment 1 point
Check bottle values
then confirm or close the
bottle and cancel
Ref.: 21821085-0/A
ULTIMA 600
Calibration in progress
Open the bottle
CO= 0,00 xvol
CO2= 0,0 xvol
HC= 0000 ppmvol
<I>
Adjustment 1 point
Close the bottle
<C>
Adjustm. 2 points
Info ajustage
Info adjustment
O2 installation
Negative value
30 s
(3,49 xvol)
(13,7 xvol)
(1020 ppmvol)
SELFTEST
5-3