Honeywell SMV800 User Manual page 163

Multivariable transmitter modbus communication
Hide thumbs Also See for SMV800:
Table of Contents

Advertisement

Bore Material
292
PV Simulation
293
Fail Safe
Simulated DP
294
Simulated SP
296
Simulated PT
298
Revision 1.0
Shows currently selected
Bore Material. This is user
configurable parameter.
Available Bore Material is
dependent upon the Flow
Calculation Standard and
Algorithm Options. Note that
only Advanced
Algorithms Options Supports
GOST material/Standard.
Configures Temperature and
Static Pressure failsafe
ON/Off conditions, Reverse
Flow ON/OFF condition and
Simulation ON / OFF
conditions for Process
Variables DP, SP, PT, Flow
Simulated DP value when
Simulation setting is ON
Simulated SP value when
Simulation setting is ON
Simulated PT value when
Simulation setting is ON
SMV800 Modbus User's Guide
R/W
uint8
When Flow Calc Standard
is other than GOST :
0 = 304 Stainless Steel
1 = 316 Stainless Steel
2 = 304/316 Stainless
Steel
3 = Corbon Steel
4 = Hastelloy
5 = Monel 400
6 = Other
When Flow Calc Standard
is GOST:
0 = 35Π
1 = 45Π
2 = 20XMΠ
3 = 12X18H9TΠ
4 = 15K,20K
5 = 22K
6 = 16ГC
7 = 09Г2C
8 = 10
9 = 15
10 = 20
11 = 30,35
12 = 40,45
13 = 10Г2
14 = 38XA
15 = 40X
16 = 15XM
17 = 30XM,30XMA
18 = 12X1MФ
19 = 25X1MФ
20 = 25X2MФ
21 = 15X5M
22 = 18X2H4MA
23 = 38XH3MФA
24 = 08X13
25 = 12X13
26 = 30X13
R/W
uint8
Bit 0 :-> 0 = Off, 1 = pv2
failsafe
Bit 1 :-> 0 = Off, 1 = pv3
failsafe
Bit 2 :-> 0 = Off, 1 =
reverse flow on
Bit 3 :-> 0 = Off, 1 =
simulate DP
Bit 4 :-> 0 = Off, 1 =
simulate SP
Bit 5 :-> 0 = Off, 1 =
simulate PT
Bit 6 :-> 0 = Off, 1 =
simulate Flow
Bit 7 :-> reserved
RO
float
RO
float
RO
float
N/A
N/A
N/A
Page 151

Advertisement

Table of Contents
loading

Table of Contents