Honeywell SMV800 User Manual page 127

Multivariable transmitter modbus communication
Hide thumbs Also See for SMV800:
Table of Contents

Advertisement

Details of
116
Critical
Fault
Revision 1.0
pres_meter
_body_failure
Bit - 0
pres_sensor
_characterizatio
Bit -1
n_corrupt
pres_suspect
_input
Bit - 2
pres_ram_DB
_Fault
Bit - 3
pres_nvm
_corrupt
Bit - 4
pres_ram
_corrupt
Bit - 5
pres_flash
Bit - 6
_CRC_Fault
pres_flow
_failure
Bit - 7
temp_input1
_fault
Bit - 8
Bit - 9
unused
SMV800 Modbus User's Guide
Pressure module is
reporting a critical failure of
the pressure sensing
measurement within the
Meter Body, which may be
caused by one of the
following:
• Meter body failure
• Sensor firmware flow
failure
Pressure module is
reporting corruption in
the Pressure
Characterization data
Differential Pressure,
Meter Body Temperature
and/orvStatic Pressure
input are extremely out of
range such that the value
is suspect.
Pressure module is
reporting corruption in
the database in the
Random-Access Memory
(RAM)
Pressure module is
reporting corruption
of the Non-Volatile Memory
data (NVM)
Pressure module is
reporting corruption in
the Random-Access
Memory (RAM)
Pressure module is
reporting corruption in
sensor firmware
Pressure module firmware
Program Flow Failure. This
fault will get set if any of
the critical part of the code
is not executed after
expected duration.
The temperature sensor
(Thermocouple or RTD)
has an open input. The
sensor connections may
be disconnected or broken.
NA
Power cycle the device. If
the problem persists,
replace the Meter Body
Power cycle the device. If
the problem persists,
replace the Meter Body
Verify that all inputs are
within specifications.
Power cycle the device. If
the problem persists,
replace the Meter Body.
Power cycle the device. If
the problem persists,
replace the Meter Body
Power cycle the device and
if fault is cleared ensure to
configure sensor
parameters. If the problem
persists, replace the Meter
Body
Power cycle the device. If
the problem persists,
replace the Meter Body
Power cycle the device. If
the problem persists,
replace the Meter Body
Power cycle the device. If
the problem persists,
replace the Meter Body
Check the temperature
sensor connections for
disconnections or broken
wires. Repair the sensor
connections
NA
Page 115