Transmitter Register Maps; Table 3-8: Smv800 Modbus Coils - Honeywell SMV800 User Manual

Multivariable transmitter modbus communication
Hide thumbs Also See for SMV800:
Table of Contents

Advertisement

3.2.12.
This section contains three register maps for SMV800 Modbus Device. These maps
include Coils, Holding register and Floating point register.
Name
Soft Reset
Differential Pressure Calib
HiPt
Differential Pressure Calib
LowPt
Static Pressure Calib_HiPt
Static Pressure Calib
LowPt
Differential Pressure Zero
Correct
Static Pressure Zero
Correct
Differential Pressure Reset
Correct
Static Pressure Reset
Correct
Process Temperature
Reset Correct
Process Temperature
Calib HiPt
Process Temperature
Calib LowPt
Critical Alarm
Warning Alarm
Differential Pressure
Above Upper Limit
Differential Pressure
Below Lower Limit
Static Pressure Above
Upper Limit
Static Pressure Below
Lower Limit
Process Temperature
Above Upper Limit
Process Temperature
Below Lower Limit
MBT Above Upper Limit
MBT Below Lower Limit
Sensor Module Failure
Sensor Communication
Failure
DP Simulation Enabled
SP Simulation Enabled
PT Simulation Enabled
Flow Simulation Enabled
WP Enabled
Page 128

Transmitter register maps

Table 3-8: SMV800 Modbus Coils

Coil
Function
Description
Address
Codes
1
5
Perform device soft reset
2
5
Perform differential pressure calibration at high point
3
5
Perform differential pressure calibration at low point
4
5
Perform static pressure calibration at high point
5
5
Perform static pressure calibration at low point
6
5
Perform differential pressure zero correct
7
5
Perform static pressure zero correct
8
5
Perform differential pressure reset correct
9
5
Perform static pressure reset correct
10
5
Perform process temperature reset corrects
11
5
Perform process temperature calibration at high point
12
5
Perform process temperature calibration at low point
49
1
Device critical alarm status
50
1
Device non-critical alarm status
54
1
Status of differential pressure above upper limit
55
1
Status of differential pressure below lower limit
60
1
Status of Static pressure above upper limit
61
1
Status of Static pressure below lower limit
68
1
Status of process temperature above upper limit
69
1
Status of process temperature below lower limit
75
1
Status of meter body temperature above Upper limit
76
1
Status of meter body temperature below lower limit
79
1
Status of pressure or temperature sensor Module failure
Status of pressure or temperature sensor communication
81
1
failure
88
1
Status of differential pressure simulation enabled
89
1
Status of static pressure simulation enabled
93
1
Status of process temperature simulation enabled
94
1
Status of flow simulation enabled
95
1
Status of write protect enabled
SMV800 Modbus User's Guide
Access
Type
W
W
W
W
W
W
W
W
W
W
W
W
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
R
Revision 1.0