Replacing The Fuses Of The Infinity Valve Drive - Agilent Technologies 1290 Infinity II User Manual

Infinitylab lc series ms flow modulator
Hide thumbs Also See for 1290 Infinity II:
Table of Contents

Advertisement

Replacing the Fuses of the Infinity Valve Drive

When
If the flow module shows no reaction.
Tools required
Description
Screwdriver
Parts required
#
2
Electrical shock
WA R N I N G
1
Switch off the instrument. The line switch is located at
the rear of the flow modulator.
MS Flow Modulator User Manual
p/n
2110-1486
Disconnect the module from line power before changing a fuse or trying to open
the hatch of the power input socket.
Never reconnect the line power before havingthe power input socket closed.

Replacing the Fuses of the Infinity Valve Drive

Description
Fuse 2 AT250 V
2
Disconnect the power cable from the power input
socket.
7
Maintenance
65