Quick Start Wizard via AMS Device Manager

Notes:
• The layout of the dialog boxes shown may vary according to the resolution setting
for your computer monitor
• At each step, you can accept the default values without modification and click on the
next step to proceed.
• After modifying parameters click on Apply inside the Quick Start window to write the
new values to the device.
• Click on OK only if you wish to update all parameters to the device and close AMS.
A Wizard provides an easy 5-step Quick Start procedure that configures the device for a
simple application.

Quick Start

Step 1 – Identification
1)
Navigate to Configure/Setup > Wizard > Quick Start.
2)
Click on Step 1.
3)
You can accept the default values without modification. (Description, Message, and
Installation Date fields can be left blank.) If desired, make changes then click on
Apply.
Step 2 – Application Type
1)
Click on Step 2.
2)
Select the application type (Level, Space, Distance, or No Service
(Liquid or Low dielectric liquid).
3)
Click on Apply.
1)
For details see
7ML19985JP02
Sensor Mode
on page 129.
SITRANS LR200 (HART) – INSTRUCTION MANUAL