Rate - Siemens Sitrans LR200 Instruction Manual

Hart
Hide thumbs Also See for Sitrans LR200:
Table of Contents

Advertisement

2.3.2. High Calibration Pt.
Distance from sensor reference point
defined in Units (2.2.1.).
Values
Related parameters
When setting the High Calibration Point value, note that echoes are ignored
within Near Range (see
2.3.3. Sensor Offset
A constant offset that can be added to or subtracted from sensor value
compensate if the sensor reference point has shifted. (For example, this could
result from adding a thicker gasket or reducing the standoff/nozzle height).
Values
Related parameters

2.4. Rate

Note: Default settings in the parameter tables are indicated with an asterisk (*)
unless explicitly stated.
2.4.1. Response Rate
Note: Changing Response Rate resets Fill Rate (2.4.2.), Empty rate (2.4.3.), and
Damping Filter (2.2.4.).
Sets the reaction speed of the device to measurement changes.
Response Rate
(2.4.1.)
*
Use a setting just faster than the maximum filling or emptying rate (whichever
is faster).
1)
The value produced by the echo processing which represents the distance from sen-
sor reference point to the target. (see Sensor Mode (2.2.2.) on page 75 for an illustra-
tion).
7ML19985JP02
SITRANS LR200 (HART) – INSTRUCTION MANUAL
Near Range (2.8.1.)
Fill Rate
(2.4.2.)
slow
0.1 m/min (0.32 ft/min)
medium
1.0 m/min (3.28 ft/min)
fast
10.0 m/min (32.8 ft/min)
1)
to High Calibration Point. Units are
Range: 0 to 20 m. Default 0.00 m
Units (2.2.1.)
Near Range (2.8. 1 .)
on page 127 for details).
Range: -99.999 to +99.999. Default: 0.00
Units (2.2.1.)
Empty rate
(2.4.3.)
1)
to
Damping Filter
(2.2.4.)
10 s
10 s
0 s
Page 77